We explore likely causes of these discrepancies and conclude that satellite losses, as recently discussed by Brundle and Crist, are a significant factor that must be accounted for if first principles RSFs are to be used. The empirically derived RSFs differ by up to 10% for several photoelectron peaks and directly impact the accuracy of the resulting measured composition. These results are then compared with standard RSF libraries provided by instrument manufacturers, as well as those derived from first principles. RSFs for three common X-ray-induced Auger lines (two Mg KLL and one Na KLL) have also been evaluated. Here, we have examined a series of in vacuo fractured silicate glass samples covering a wide range of elemental compositions that allow determination of empirical RSFs for the photoelectron peaks of several important elements. Understanding of the key factors that impact the amount of signal observed from a given photoelectron peak has improved substantially over the past 30+ years, which permits calculation of RSFs from first principles. ![]() Quantification of spectral data from X-ray photoelectron spectroscopy is frequently done using relative sensitivity factors (RSFs) provided by instrument or software vendors.